The sequence below separates what the circuit must do before digitization from what software can do afterward. Specific hardware and implementation choices are still subject to prototyping and verification.
01 / BIOELECTRIC INTERFACECapture body-surface potentials
RA, LA, and LL are limb-designated sensing sites positioned on the torso; V is one movable chest sensing site. RL is intended as a bias/right-leg-drive connection, if supported by the selected AFE.
V_lead II = V_LL − V_RAKey risk: contact impedance, poor placement, displacement, motion, and common-mode pickup.
02 / ANALOG PROTECTIONPreserve microvolt information
Patient-input protection, an ECG-grade amplifier, common-mode control, and analog anti-alias conditioning must be designed before the ADC. Digital filters cannot restore information lost to clipping, saturation, or aliasing.
Design questions: AFE noise, CMRR, gain range, input protection, and lead-off detection.
03 / ACQUISITIONSample simultaneously
The design target is at least 500 samples/s per acquired channel, with at least 16-bit conversion resolution. Compare actual simultaneous sample timing, noise, and effective resolution once the AFE is chosen.
t[k] = k / fₛ ; fₛ = 500 Hz target
04 / DIGITAL CONDITIONINGSuppress noise with care
Estimate and attenuate baseline wander, mains interference, and excess high-frequency noise. A narrow rhythm display may be easier to read, but its output must not be presented as undistorted diagnostic morphology.
Important: causal real-time high-pass filters can shift ST features; zero-phase forward–reverse filtering needs future samples.
05 / LEAD MATH + FEATURESCompute views and status
Use synchronized sensing potentials to form limb and chest lead views. Additional computations may include QRS detection, heart rate, signal-quality indicators, lead-off flags, and battery/network status.
III = II − I ; WCT = (RA+LA+LL)/3
06 / TELEMETRY + ALARMSDeliver traceable packets
Attach sequence numbers, acquisition timestamps, channel identities, quality and fault flags, then transmit over authenticated protected Wi-Fi to the receiving station.
Failure mode: packet loss must appear as missing samples or a connectivity alert—not be silently smoothed into plausible ECG.