Engineering deep dive · EasyG telemetry

From microvolt signals to meaningful ECG telemetry.

Explore the proposed digital signal processing (DSP) path inside EasyG: how small voltage differences are acquired, protected, digitized, conditioned, transformed into ECG lead views, checked for faults, and delivered to a hospital monitor.

DESIGN EXPLAINER · INTERACTIVE DATA ARE SYNTHETIC
Signal chain overview

One continuous path. Multiple engineering safeguards.

01 · Electrode interfaceRA / LA / LL / V + RL
→
02 · Analog front endProtection, gain, bias, lead-off
→
03 · ADC + timestampsSample signals simultaneously
→
04 · Digital conditioningFilter, assess quality
→
05 · Derived ECGLead equations + heart rate
→
06 · Wi-Fi telemetryPacketize, transmit, alarm
System boundary: EasyG currently proposes a centralized patient-worn transmitter with short electrode leads, not the earlier UWB-synchronized distributed-node architecture. The ESP32 handles digital processing and communications; a dedicated ECG analog front end is needed to measure the small signals safely and accurately.
Detailed walkthrough

What happens at every stage?

The sequence below separates what the circuit must do before digitization from what software can do afterward. Specific hardware and implementation choices are still subject to prototyping and verification.

01 / BIOELECTRIC INTERFACE

Capture body-surface potentials

RA, LA, and LL are limb-designated sensing sites positioned on the torso; V is one movable chest sensing site. RL is intended as a bias/right-leg-drive connection, if supported by the selected AFE.

V_lead II = V_LL − V_RA

Key risk: contact impedance, poor placement, displacement, motion, and common-mode pickup.

02 / ANALOG PROTECTION

Preserve microvolt information

Patient-input protection, an ECG-grade amplifier, common-mode control, and analog anti-alias conditioning must be designed before the ADC. Digital filters cannot restore information lost to clipping, saturation, or aliasing.

Design questions: AFE noise, CMRR, gain range, input protection, and lead-off detection.

03 / ACQUISITION

Sample simultaneously

The design target is at least 500 samples/s per acquired channel, with at least 16-bit conversion resolution. Compare actual simultaneous sample timing, noise, and effective resolution once the AFE is chosen.

t[k] = k / fₛ   ;   fₛ = 500 Hz target
04 / DIGITAL CONDITIONING

Suppress noise with care

Estimate and attenuate baseline wander, mains interference, and excess high-frequency noise. A narrow rhythm display may be easier to read, but its output must not be presented as undistorted diagnostic morphology.

Important: causal real-time high-pass filters can shift ST features; zero-phase forward–reverse filtering needs future samples.

05 / LEAD MATH + FEATURES

Compute views and status

Use synchronized sensing potentials to form limb and chest lead views. Additional computations may include QRS detection, heart rate, signal-quality indicators, lead-off flags, and battery/network status.

III = II − I   ;   WCT = (RA+LA+LL)/3
06 / TELEMETRY + ALARMS

Deliver traceable packets

Attach sequence numbers, acquisition timestamps, channel identities, quality and fault flags, then transmit over authenticated protected Wi-Fi to the receiving station.

Failure mode: packet loss must appear as missing samples or a connectivity alert—not be silently smoothed into plausible ECG.

Interactive educational simulation

Signal Processing Laboratory

Toggle realistic classes of interference and compare the original synthetic voltage difference with the output of two illustrative first-order digital filter stages. All seven displayed lead views are calculated from a shared set of synthetic electrode potentials. This is not an AFE emulator or clinical-grade DSP.

EasyG virtual oscilloscope SYNTHETIC ONLY

500 simulated samples/second · derived potential differences · no hardware connected · demonstration filters are causal single-pole approximations
STATUS: SYNTHETIC DATACHECK: II − I − III ≈ 0No hardware, patient or clinical diagnosis
1 · RAW SYNTHETIC DIFFERENTIAL VOLTAGEPre-digital-conditioning · mV
2 · FILTERED OUTPUTIllustrative response · mV
3 · FILTER DISTORTION / DIFFERENCEfiltered minus raw · mV

Plots use a fixed-length synthetic recording, regenerated deterministically on control changes. The first-order filters are implemented for education; the nominal cutoff labels do not establish AHA compliance, ST fidelity, or a validated biomedical filter. A lead-off condition suppresses affected views rather than inventing a replacement waveform.

Reference geometry

How four sensing sites become seven displayed views

Frontal-plane leads

Three limb-designated electrodes provide two independent frontal-plane differences; the remaining limb lead views are algebraic combinations.

I = LA − RA
II = LL − RA
III = LL − LA = II − I
aVR = RA − (LA + LL)/2
aVL = LA − (RA + LL)/2
aVF = LL − (RA + LA)/2

The exact voltages depend on anatomical electrode placement. Torso-based limb recordings are modified leads and should not be labeled equivalent to standard limb-electrode ECG without validation.

Wilson Central Terminal (WCT)

WCT is a calculated voltage reference made by averaging the RA, LA, and LL electrode potentials. It is neither a separate skin electrode nor a physical electrical ground.

WCT = (RA + LA + LL) / 3
V = V_chest − WCT

The one movable V electrode provides one chest view at a time, not V1–V6 simultaneously. WCT is distinct from the active RL bias/right-leg-drive circuit.

Source: Kligfield et al., AHA/ACCF/HRS ECG standardization statement (2007), DOI: 10.1016/j.jacc.2007.01.024.

Why filtering is a clinical question

Acquisition bandwidth is not the same as display bandwidth

Wider-bandwidth acquisition

The AHA/ACCF/HRS 2007 statement supports an adult upper-frequency response of at least 150 Hz for diagnostic ECG and describes 500 samples/s as an appropriate acquisition target under its stated error assumptions.

A 0.05 Hz low-frequency cutoff is traditionally used to preserve ST segments. Certain higher digital cutoffs are discussed only with phase-distortion qualifications.

Rhythm monitoring display

A narrower display, such as 0.5–40 Hz, can suppress drift and high-frequency noise and make rhythm interpretation easier. But it can change amplitudes and morphology, so it must not silently substitute for a diagnostic-bandwidth representation.

Bidirectional, zero-phase filtering needs future samples and introduces delay; it cannot be treated as instantaneously causal bedside processing.

Engineering caution: the simulator uses simplified first-order filters only to visualize signal tradeoffs. Real EasyG filter design must be evaluated for frequency response, group delay, settling, ST-level error, QRS amplitude error, and compatibility with alarm latency. AHA/ACCF/HRS recommendations are scientific consensus statements, not proof of regulatory compliance.
Full-resolution visual references

ECG posters

These educational illustrations provide a broader view of ECG physiology and electronics. Open the original PNGs to inspect small text and diagrams without reducing image resolution.

ECG concept map poster, full resolution
Download ECG concept map · 1536 × 1024
Electrical engineering behind electrocardiograms infographic, full resolution
Download ECG engineering infographic · 1055 × 1491
From demonstration to evidence

A verification plan tied to measurable failure modes

LayerVerifyMethod and evidenceCurrent state
Electrical fidelityNoise, gain, CMRR, bandwidth, saturation and patient-input safetyCalibrated ECG simulator, swept-frequency input, shorted-input noise and safety assessmentNot yet reported
Sampling + timingPer-channel sample rate, acquisition skew, missing samplesKnown periodic test signals and timestamps; quantify interchannel timing errorNot yet reported
Computational fidelityDerived-lead identities, filter distortion, QRS/HR outputsUnit tests including II = I + III; synthetic and recorded-reference datasetsDemo only
Fault handlingLead-off, lead swaps, battery status, Wi-Fi disconnectsControlled disconnection/reconnection and packet-loss injection with timed alarm logsNot yet reported
Clinical agreementWaveform amplitudes, durations, patient movement, electrode placementAppropriately approved comparison study against simultaneous reference monitoring, across relevant body morphologiesFuture work
Do not confuse three kinds of success: correct digital arithmetic does not prove analog fidelity; a clean bench signal does not prove clinical agreement; and similarity to a reference waveform does not establish diagnostic equivalence across patient populations.
Reading and traceability

References and further engineering work

  1. Kligfield P, et al. Recommendations for the Standardization and Interpretation of the Electrocardiogram, Part I: The Electrocardiogram and Its Technology. J Am Coll Cardiol. 2007;49(10):1109–1127. DOI. Used for acquisition bandwidth, sampling, timing, lead relationships, modified torso-lead caveats, and validation context.
  2. Lv B, et al. 3-Node UWB-synchronized wearable system for lead-wire-free 12-lead ECG derivation. Measurement. 2026;282:122044. DOI. An earlier related research architecture, not the current EasyG transmitter or EasyG test result.
  3. PTB-XL: A large publicly available electrocardiography dataset. PhysioNet. Candidate resource for offline analysis; not used by this browser simulation.

Future implementation must replace the educational synthetic-data generator and placeholder filter design with a documented acquisition driver, verified DSP and telemetry protocol. Clinical decision support and alarm use require additional validation and safeguards.