ENGINEERING CONCEPT • SYNTHETIC ECG DEMONSTRATION • NO LIVE PATIENT DATA • NOT FOR CLINICAL USE
Arizona State University • Biomedical Engineering • Capstone Team #29

EasyG: Five-Electrode Wireless ECG

A capstone design for wireless hospital telemetry that aims to give patients more freedom to move.

Capstone Engineering Team (#29)
Brianna Branson Kyle Cagno Jay Evans Gabriel Baca
Faculty Mentor Dr. Xiao Wang Arizona State University — Ira A. Fulton Schools of Engineering, School of Biological and Health Systems Engineering
Clinical Mentors
Dr. Sara Belko The University of Arizona
Dr. Srekar Ravi University of Maryland

The demos run in your browser with synthetic ECG signals. Hardware verification is planned.

EASYG-TX01 • LEAD II STREAM
DEMO · OFFLINE
Synthetic ECG · animated preview
0.05–150 Hz • Notch Filtered
72 BPM
SIMULATED NORMAL RHYTHM
≥ 500 Samples/s
Per Channel
≥ 16-Bit ADC
Resolution Target
5 Electrodes → 7 Views
6 limb + 1 chest · proposed
01 • Clinical Need & Engineering Targets

Wires Lead to Delay in Urgent Situations

Clinical Problem & Bedside Tether Burden

Conventional bedside electrocardiogram (ECG/EKG) telemetry enables continuous cardiac monitoring for hospitalized patients, but standard systems burden patients with long lead wires and bulky devices that restrict mobility, pull on electrodes, and complicate routine hospital workflow.

EasyG proposes a compact patient-worn transmitter to support continuous hospital ECG telemetry without a bedside monitor tether, with signal fidelity and safe operation as engineering targets that remain to be validated.

Designed to Remove Bedside Tether Aims to Reduce Wire Snags Designed for Easier Patient Access
Team #29 Mission Statement
“Our team aims to design a five-electrode wearable ECG telemetry system that supports greater inpatient mobility and clinician access by replacing the long bedside tether with a wireless connection.”
INPATIENT TELEMETRY 5-LEAD WIRELESS

Preliminary Engineering Targets

BASELINE DESIGN SPECIFICATIONS
01
ECG Sampling
≥ 500 samples/s

Per acquired channel for diagnostic QRS and rhythm fidelity.

02
ADC Resolution
≥ 16-bit

High-dynamic-range digitization preserving microvolt features.

03
Mobility
Short Leads & Pod

Compact wearable sternum transmitter with zero bedside tether.

04
Reliability
Active Fault Flags

Real-time lead-off, battery status, and Wi-Fi link-loss alerts.

05
Clinical Workflow
Rapid Setup

Simple placement, intuitive pairing, and hospital-grade cleaning.

02 • Proposed Patient-Worn Hardware

Interactive Five-Electrode Wearable Transmitter

Click the central sternum pod or any of the 5 coiled-lead electrodes (RA, LA, V, RL, LL) to inspect its anatomical placement and electrical function.

ANTERIOR THORAX VIEW • X-COIL ARCHITECTURE INTERACTIVE DIAGRAM
RA LA V RL LL EasyG TX-5L PATIENT RIGHT PATIENT LEFT
CENTRAL TRANSMITTER POD

Sternum-Mounted Wireless Pod

AFE + ESP32 + WI-FI

Compact, lightweight patient-worn enclosure positioned over the mid-sternum. Houses input protection, multi-channel Analog Front End (AFE), ≥16-bit ADC, ESP32 MCU, protected Wi-Fi radio, and rechargeable battery—while eliminating the long wired link to the bedside monitor in the proposed architecture.

Anatomical Placement
Mid-sternum (central anterior chest wall) with medical-grade skin adhesive
Electrical & Lead Role
Proposed RA, LA, LL and V sensing plus RL bias; six limb leads and one precordial view can be calculated from supported input potentials
Ergonomic & Clinical Benefit
Short coiled leads prevent snagging during patient turns while preserving unobstructed lateral & apical access
Engineering evidence · AHA/ACCF/HRS 2007

How acquisition technology changes clinical interpretation

The AHA/ACCF/HRS scientific statement (not an evidence-graded guideline) explains that electrode placement, acquisition bandwidth, filter phase response, timing and software can alter ECG amplitudes, intervals and resulting interpretations. EasyG uses these recommendations to define design targets and tests—not claims of measured performance or regulatory compliance. Read Kligfield et al. (2007) ↗

AHA evidence translated into EasyG proposed engineering targets
Design variable2007 AHA scientific statementEasyG proposed action / status
Sampling500 samples/s supports an adult 150 Hz upper cutoff with controlled amplitude error (pp. 1309–1310).Target ≥500 samples/s per acquired channel; measure actual sample rate and timing. Unverified.
Upper bandwidthAt least 150 Hz for adult/adolescent diagnostic acquisition; 250 Hz for pediatric considerations (pp. 1309–1310).Evaluate 150 Hz adult diagnostic-quality pathway where supported by the AFE, anti-aliasing and sampling design. Not yet demonstrated.
Low-frequency response0.05 Hz routine cutoff; up to 0.67 Hz with suitable zero-phase digital filtering, subject to phase/ST fidelity (p. 1309).Specify and characterize causal real-time filtering separately; measure ST deviation and latency before choosing a monitoring setting.
Lead alignmentSimultaneous channels should remain aligned within 10 ms, ideally better (p. 1315).Test interchannel acquisition alignment and packet timestamps independently; Wi-Fi packet arrival is not the sampling clock.
Faults / qualityFlag high electrode impedance and detect possible lead swaps (p. 1318).Verify lead-off alerts, recovery, and lead-swap screening if implemented; record false alerts and detection delay.

Source: Kligfield P, et al. Recommendations for the Standardization and Interpretation of the Electrocardiogram: Part I: The Electrocardiogram and Its Technology. J Am Coll Cardiol. 2007;49(10):1109–1127. DOI ↗. Page numbers above follow the Circulation-version pagination provided in the project reading notes (pp. 1307–1320).

03 • Hardware & Telemetry Architecture

End-to-End System Architecture & DSP Pipeline

Click any subsystem block below to inspect its hardware interfaces, signal conditioning role, and clinical telemetry safeguards.

TIER 1

On-Body Signal Acquisition

PATIENT-WORN TRANSMITTER POD
PROTECTED HOSPITAL WI-FI TELEMETRY LINK
TIER 2

Hospital Network / Monitoring

CLINICAL INFRASTRUCTURE & CENTRAL STATION
TIER 1 • ON-BODY SIGNAL ACQUISITION

5 Electrodes (RA, LA, LL, V, RL)

BIOELECTRIC INTERFACE
Inputs

Millivolt-level cardiac surface potentials across 4 sensing electrodes + RL common-mode bias

Core Processing Function

Couples ionic skin potentials into low-impedance Ag/AgCl gel contacts via short coiled leads

Outputs & Safeguards

Sensing inputs and bias connection; actual AFE channel topology and lead-off method to be established

SIGNAL PROCESSING PIPELINE

5-Stage On-Body Digital Signal Processing Flow

REAL-TIME EMBEDDED DSP
01

Acquire

Sample electrode potentials simultaneously at ≥ 500 SPS with ≥ 16-bit resolution.

02

Condition

Reduce baseline drift + powerline/motion interference while preserving diagnostic ST-T morphology.

03

Compute

Calculate derived leads (I, II, III, aVR, aVL, aVF, V) + simulated heart rate.

04

Package

Frame timestamped ECG data + signal quality index, lead-off flags, and battery status.

05

Transmit

Stream telemetry packets securely over protected hospital Wi-Fi to the floor gateway.

Explore the full signal processing pipeline

Trace electrode signals through the analog front end, sampling, digital filters, lead derivation, fault detection, and Wi-Fi telemetry—with an interactive synthetic filtering laboratory.

Open the Signal Processing Deep Dive →

Try the Wi-Fi ECG Telemetry Workstation

A browser-based monitoring workstation for the ESP32 + ADS1294 three-lead concept: live DSP controls, heart-rate and R-R readouts, and a 500 SPS synthetic ECG generator so you can explore it without hardware.

Open the Live Workstation →
04 • Cardiac Vector Math & Synthetic Waveform Explorer

ECG Lead Relationships & Einthoven’s Triangle

Δ

Einthoven’s Triangle: “Einthoven's triangle is an imaginary equilateral triangle formed by electrodes on the right arm, left arm, and left leg, representing the three standard bipolar ECG leads (I, II, and III). Each lead measures the electrical potential difference between two electrodes, allowing clinicians to observe the heart's electrical activity from three different directions in the frontal plane.”

SELECT A DISPLAYED VIEW (6 LIMB + 1 CHEST, ALL SYNTHETIC)

Interactive Lead Vector & Morphology Selector

TELEMETRY STATE:
Bipolar Limb Leads
Augmented Limb Leads
Precordial Lead (WCT reference)
Frontal Plane Vector Axis AXIS: +60°
WCT − + RA LA LL V (Chest)
Selected Lead Vector WCT = (V_RA + V_LA + V_LL)/3
BIPOLAR LIMB LEAD

Lead II

Lead II = V_LL − V_RA

Measures potential difference from Right Arm (−) to Left Leg (+) along the +60° frontal axis—closely aligned with the heart’s mean depolarization vector, yielding prominent upright P, R, and T waves.

SIMULATION: NORMAL RHYTHM • SYNTHETIC ELECTRODES CONNECTED
SIMULATION · NOT DEVICE DATA

SYNTHETIC TRAINING SIGNAL • SHARED ELECTRODE POTENTIAL MODEL • NO PATIENT DATA

GAIN: 10 mm/mV SWEEP: 25 mm/s FILTER: 0.05–150 Hz
HR: 72 BPM
Lead geometry · what seven views do and do not mean

Five electrodes ≠ seven independent channels

EasyG proposes three torso-mounted limb-sensing sites (RA, LA, LL), one movable precordial site (V), and a bias/right-leg-drive site (RL). The first three provide two independent limb-lead differences; the other four limb views are calculated. A single V electrode adds one chest view at its chosen position, rather than six simultaneous V1–V6 recordings.

Measured information

Two independent frontal-plane voltage differences plus one chest-versus-Wilson-reference difference in the intended configuration. The precise analog channel implementation depends on the selected AFE.

Derived display views

Six conventional limb-lead relationships can be computed algebraically; for example III = II − I. The website's educational simulator enforces these relationships with shared synthetic potentials.

Not interchangeable with 12-lead ECG

Torso limb placement (Mason–Likar-type modifications) changes recorded waveforms; the statement says they cannot simply be treated as standard limb-placement tracings or compared serially as equivalent (p. 1316).

Labeling rule for any future results: “Modified torso-electrode ECG — derived limb views; one measured-position precordial view.” Do not present the output as a conventional 12-lead or claim diagnostic equivalence. The 2007 statement also cautions that reduced-lead synthesis cannot duplicate conventional tracings and requires explicit labeling (pp. 1316–1317). Source ↗

Placement and inclusive wearability testing

Body morphology, breast tissue, skin characteristics, adhesive contact, motion and clinical access can affect signal quality and usability. Proposed testing should document placement landmarks, adaptations across body types, movement artifacts and any systematic differences in signal quality or adhesion. This is an EasyG design recommendation informed by the statement's placement-error discussion (pp. 1313, 1318), not a claim that the 2007 paper validated EasyG's anatomy-specific approach.

05 • Design Controls & Planned Benchtop Validation

Core Design Requirements

Five foundational engineering pillars governing hardware, firmware, and hospital telemetry integration.

Signal Fidelity

Preserve interpretable ECG morphology across bipolar, augmented, and precordial views.

Mobility

Minimize weight, bulk, and snag hazards with short coiled leads and a compact sternum pod.

Reliability

Continuously indicate signal quality, electrode lead-off, battery level, and wireless link loss.

Workflow

Enable rapid anatomical placement, quick patient pairing, and streamlined hospital disinfection.

Safety & Privacy

Ensure patient electrical isolation/protection and encrypted hospital wireless data transfer.

Proposed Verification Plan

Structured benchtop verification protocols mapped to each clinical and engineering requirement.

PLANNED BENCHTOP TESTING — RESULTS PENDING
ECG FIDELITY PENDING

Compare simulated/reference waveforms, amplitude, and morphology against calibrated ECG simulator outputs.

TIMING PENDING

Measure sample rate consistency (≥ 500 SPS), packet timestamps, and dropped samples under sustained streaming.

MOTION ARTIFACTS PENDING

Assess signal quality index and baseline stability during simulated patient movement and coiled-lead handling.

LEAD-OFF + ALARMS PENDING

Confirm immediate detection and central-display indication of disconnected or high-impedance electrodes.

WIRELESS RESILIENCE PENDING

Measure link loss recovery, access-point roaming reconnection, and fail-safe patient ID association.

POWER + WEARABILITY PENDING

Evaluate continuous battery runtime, skin adhesive retention, and ergonomic comfort across body types.

Verification traceability · no measured EasyG performance yet

Three separate kinds of fidelity

A visually plausible simulated waveform is not proof that the electronics, lead calculations or clinical decisions are accurate. These three verification tracks require different evidence.

01 · ELECTRICAL

Acquisition fidelity

Inject calibrated ECG simulator signals into the AFE. Record measured bandwidth, input noise, gain/amplitude error, sample rate, channel alignment and filter-induced ST deviation. Log instrument calibration and test conditions.

STATUS: PLANNED — no bench results supplied
02 · COMPUTATIONAL

Derived-lead fidelity

Confirm identities such as II = I + III under a shared electrode-potential model, then quantify latency, numerical error and waveform changes under noise, lead-off, lead swapping and motion artifacts. Compare filtered versus unfiltered features.

STATUS: SYNTHETIC DEMO — not hardware verification
03 · CLINICAL

Reference agreement

For future appropriately approved studies, compare EasyG with a reference ECG under matched posture, timing and placement documentation. Report waveform amplitudes, intervals, error distributions and pathology-relevant failure modes separately.

STATUS: NOT STARTED — no clinical equivalence claim

Measurement and labeling safeguards

If representative/median beats are constructed, document the algorithm and resulting noise; the AHA statement notes approximately 5 µV template noise for resolving 20 µV deflections to 10% error (pp. 1310–1311). Use time-coherent multi-lead measurements for global intervals when available; do not infer global durations from one display view. Label simulated, derived, modified-placement and recorded data distinctly. Physician interpretation and independent validation would be required before diagnostic claims.

Design rationale: Kligfield et al., AHA/ACCF/HRS scientific statement (2007) ↗. Test plans are proposed EasyG work, not completed results.

07 · The design journey

Why the architecture changed

EasyG began with exploratory research into distributed, wire-free ECG acquisition and mathematical lead reconstruction. Concept selection narrowed the scope to a five-electrode, single-transmitter hospital telemetry system. The current concept retains short electrode wires while replacing the long connection to bedside monitoring equipment with a proposed Wi-Fi link.

Illustrated engineering concepts considered: distributed nodes, flexible electrode strip, wireless electrode clips, and centralized wearable transmitter
Design exploration: multiple possible wearable ECG architectures, not four validated devices.
Concept illustration showing chest electrodes leading to a patient-worn transmitter and wireless monitoring workflow
EasyG's selected direction: a compact patient-worn transmitter with short electrode leads. Concept art, not a photograph of a working prototype.
Conceptual five-electrode placement layout with RA, LA, LL, RL and movable precordial V electrode
Electrode layout reference (conceptual). Exact positioning and patient-specific fit require clinical and engineering review.
01 · RESEARCH

Distributed ECG nodes

Investigated UWB timing, local differential acquisition, electrode geometry and methods that derive conventional leads from a reduced measurement set.

02 · CONCEPT SELECTION

Wearability tradeoffs

Compared distributed nodes, flexible strips, transmitter clips and a centralized wearable using clinical usability, complexity, signal quality, cost and engineering risk.

03 · CURRENT DIRECTION

Hospital telemetry first

Focused on dependable ECG acquisition and wireless transport through a patient-worn ESP32 system, with measured performance and clinical workflow integration still pending.

Evidence boundary

Published feasibility is not EasyG validation.

Lv and colleagues (2026) reported a distinct three-node UWB-synchronized lead-derivation system. Their reported synchronization and waveform agreement outcomes describe their experimental system, not this capstone prototype. EasyG's present five-electrode Wi-Fi design uses a different architecture and must be characterized independently.

08 · Virtual benchtop laboratory

From simulation to measured evidence

The interactive ECG explorer above uses a single set of synthetic electrode potentials to calculate its displayed lead views. This is an educational mathematical model—not a test of an actual ECG analog front end, medical alarm, or Wi-Fi telemetry link.

Conceptual verification process from simulated ECG input through sensing hardware, wireless communication and waveform analysis
Proposed benchtop test workflow; no measured performance values have been supplied.
Conceptual block diagram of electrodes, ECG analog front end, ESP32 wearable transmitter, Wi-Fi and monitoring station
Architecture sketch; AFE selection, circuit-level protection, isolation approach and network implementation are still design tasks.
1 · SyntheticCompute electrode potentials and consistent displayed leads.
2 · Reference ECGLoad de-identified waveform records and annotate provenance.
3 · Bench acquisitionCollect ECG-simulator signals from the selected AFE.
4 · Wireless telemetryQuantify timestamps, packet loss, latency and recovery.
5 · VerificationReport observed results, uncertainties and acceptance criteria.
10 · Full-resolution educational posters

ECG concept map & electrical engineering infographic

Select a poster to open the original PNG at native resolution. The diagrams are educational reference illustrations; some depict general ECG systems or earlier multi-node design concepts rather than the selected EasyG prototype.

Detailed ECG concept map connecting cardiac electrophysiology, signal generation, biophysics, equations, signal characteristics, requirements, processing and architecture
ECG concept map1536 × 1024 pixels · full original resolutionDownload original PNG
Educational ECG engineering infographic showing cardiac electrical origins, 12 lead placement, skin electrode interface, front end, filtering, analog to digital conversion, and signal specifications
Electrical engineering behind electrocardiograms1055 × 1491 pixels · full original resolutionDownload original PNG
09 · Literature and data

Research basis and sources

These references inform the project design and future validation. None should be interpreted as experimental verification of EasyG's current architecture.

FOUNDATIONAL SCIENTIFIC STATEMENT · 2007

AHA/ACCF/HRS ECG technology and standardization

Kligfield P, et al. Recommendations for the Standardization and Interpretation of the Electrocardiogram: Part I: The Electrocardiogram and Its Technology. J Am Coll Cardiol. 2007;49(10):1109–1127. A scientific statement, not an evidence-graded guideline. Supports acquisition, filtering, lead-placement and validation rationale; it does not establish EasyG compliance.

Read the statement ↗
PUBLISHED RESEARCH

Three-node UWB ECG derivation (2026)

B. Lv et al., Measurement 282, 122044. Separate three-node architecture using synchronized acquisition and session-specific lead transformation.

Read published article ↗
PUBLIC DATASET

PTB-XL ECG database

Large public clinical ECG dataset that may support offline algorithm exploration; it is not a substitute for bench characterization of the EasyG acquisition hardware.

Explore PTB-XL ↗
DESIGN REFERENCE

IEC 60601-2-27

Particular requirements concerning the basic safety and essential performance of electrocardiographic monitoring equipment. Compliance has not been assessed.

FUTURE VERIFICATION

Prototype evidence log

The site will distinguish numerical design targets from independently measured results, test conditions, instrument calibration and documented pass/fail decisions.

06 • Next Milestones & Regulatory Basis

Future Directions Roadmap

PHASE 01 HARDWARE

Finalize BOM & Compact PCB / Analog Front End

Lock component selection for low-noise AFE, ≥16-bit ADC, ESP32 wireless module, and miniaturized enclosure footprint.

PHASE 02 BENCHTOP

Characterize Noise, Battery Life, Wi-Fi Recovery & Fault Alarms

Quantify input-referred noise floor, continuous shift runtime, network drop resilience, and lead-off alarm latency.

PHASE 03 INTEGRATION

Optimize Device Adhesion & Infrastructure Compatibility

Refine skin-contact adhesive ergonomics and validate seamless compatibility with managed hospital Wi-Fi gateways.

PHASE 04 VALIDATION

Verify Safety & Performance

Execute comprehensive electrical safety, cybersecurity, and diagnostic waveform verification against clinical benchmarks.

REFERENCES / DESIGN BASIS

Clinical & Regulatory Standards Alignment

AHA hospital ECG monitoring practice standards IEC 60601-2-27 FDA wireless medical-device guidance