Acquisition fidelity
Inject calibrated ECG simulator signals into the AFE. Record measured bandwidth, input noise, gain/amplitude error, sample rate, channel alignment and filter-induced ST deviation. Log instrument calibration and test conditions.
A capstone design for wireless hospital telemetry that aims to give patients more freedom to move.
The demos run in your browser with synthetic ECG signals. Hardware verification is planned.
Conventional bedside electrocardiogram (ECG/EKG) telemetry enables continuous cardiac monitoring for hospitalized patients, but standard systems burden patients with long lead wires and bulky devices that restrict mobility, pull on electrodes, and complicate routine hospital workflow.
EasyG proposes a compact patient-worn transmitter to support continuous hospital ECG telemetry without a bedside monitor tether, with signal fidelity and safe operation as engineering targets that remain to be validated.
“Our team aims to design a five-electrode wearable ECG telemetry system that supports greater inpatient mobility and clinician access by replacing the long bedside tether with a wireless connection.”
Per acquired channel for diagnostic QRS and rhythm fidelity.
High-dynamic-range digitization preserving microvolt features.
Compact wearable sternum transmitter with zero bedside tether.
Real-time lead-off, battery status, and Wi-Fi link-loss alerts.
Simple placement, intuitive pairing, and hospital-grade cleaning.
Click the central sternum pod or any of the 5 coiled-lead electrodes (RA, LA, V, RL, LL) to inspect its anatomical placement and electrical function.
Compact, lightweight patient-worn enclosure positioned over the mid-sternum. Houses input protection, multi-channel Analog Front End (AFE), ≥16-bit ADC, ESP32 MCU, protected Wi-Fi radio, and rechargeable battery—while eliminating the long wired link to the bedside monitor in the proposed architecture.
The AHA/ACCF/HRS scientific statement (not an evidence-graded guideline) explains that electrode placement, acquisition bandwidth, filter phase response, timing and software can alter ECG amplitudes, intervals and resulting interpretations. EasyG uses these recommendations to define design targets and tests—not claims of measured performance or regulatory compliance. Read Kligfield et al. (2007) ↗
| Design variable | 2007 AHA scientific statement | EasyG proposed action / status |
|---|---|---|
| Sampling | 500 samples/s supports an adult 150 Hz upper cutoff with controlled amplitude error (pp. 1309–1310). | Target ≥500 samples/s per acquired channel; measure actual sample rate and timing. Unverified. |
| Upper bandwidth | At least 150 Hz for adult/adolescent diagnostic acquisition; 250 Hz for pediatric considerations (pp. 1309–1310). | Evaluate 150 Hz adult diagnostic-quality pathway where supported by the AFE, anti-aliasing and sampling design. Not yet demonstrated. |
| Low-frequency response | 0.05 Hz routine cutoff; up to 0.67 Hz with suitable zero-phase digital filtering, subject to phase/ST fidelity (p. 1309). | Specify and characterize causal real-time filtering separately; measure ST deviation and latency before choosing a monitoring setting. |
| Lead alignment | Simultaneous channels should remain aligned within 10 ms, ideally better (p. 1315). | Test interchannel acquisition alignment and packet timestamps independently; Wi-Fi packet arrival is not the sampling clock. |
| Faults / quality | Flag high electrode impedance and detect possible lead swaps (p. 1318). | Verify lead-off alerts, recovery, and lead-swap screening if implemented; record false alerts and detection delay. |
Source: Kligfield P, et al. Recommendations for the Standardization and Interpretation of the Electrocardiogram: Part I: The Electrocardiogram and Its Technology. J Am Coll Cardiol. 2007;49(10):1109–1127. DOI ↗. Page numbers above follow the Circulation-version pagination provided in the project reading notes (pp. 1307–1320).
Click any subsystem block below to inspect its hardware interfaces, signal conditioning role, and clinical telemetry safeguards.
Millivolt-level cardiac surface potentials across 4 sensing electrodes + RL common-mode bias
Couples ionic skin potentials into low-impedance Ag/AgCl gel contacts via short coiled leads
Sensing inputs and bias connection; actual AFE channel topology and lead-off method to be established
Sample electrode potentials simultaneously at ≥ 500 SPS with ≥ 16-bit resolution.
Reduce baseline drift + powerline/motion interference while preserving diagnostic ST-T morphology.
Calculate derived leads (I, II, III, aVR, aVL, aVF, V) + simulated heart rate.
Frame timestamped ECG data + signal quality index, lead-off flags, and battery status.
Stream telemetry packets securely over protected hospital Wi-Fi to the floor gateway.
Trace electrode signals through the analog front end, sampling, digital filters, lead derivation, fault detection, and Wi-Fi telemetry—with an interactive synthetic filtering laboratory.
Open the Signal Processing Deep Dive →A browser-based monitoring workstation for the ESP32 + ADS1294 three-lead concept: live DSP controls, heart-rate and R-R readouts, and a 500 SPS synthetic ECG generator so you can explore it without hardware.
Open the Live Workstation →Einthoven’s Triangle: “Einthoven's triangle is an imaginary equilateral triangle formed by electrodes on the right arm, left arm, and left leg, representing the three standard bipolar ECG leads (I, II, and III). Each lead measures the electrical potential difference between two electrodes, allowing clinicians to observe the heart's electrical activity from three different directions in the frontal plane.”
Measures potential difference from Right Arm (−) to Left Leg (+) along the +60° frontal axis—closely aligned with the heart’s mean depolarization vector, yielding prominent upright P, R, and T waves.
SYNTHETIC TRAINING SIGNAL • SHARED ELECTRODE POTENTIAL MODEL • NO PATIENT DATA
EasyG proposes three torso-mounted limb-sensing sites (RA, LA, LL), one movable precordial site (V), and a bias/right-leg-drive site (RL). The first three provide two independent limb-lead differences; the other four limb views are calculated. A single V electrode adds one chest view at its chosen position, rather than six simultaneous V1–V6 recordings.
Two independent frontal-plane voltage differences plus one chest-versus-Wilson-reference difference in the intended configuration. The precise analog channel implementation depends on the selected AFE.
Six conventional limb-lead relationships can be computed algebraically; for example III = II − I. The website's educational simulator enforces these relationships with shared synthetic potentials.
Torso limb placement (Mason–Likar-type modifications) changes recorded waveforms; the statement says they cannot simply be treated as standard limb-placement tracings or compared serially as equivalent (p. 1316).
Labeling rule for any future results: “Modified torso-electrode ECG — derived limb views; one measured-position precordial view.” Do not present the output as a conventional 12-lead or claim diagnostic equivalence. The 2007 statement also cautions that reduced-lead synthesis cannot duplicate conventional tracings and requires explicit labeling (pp. 1316–1317). Source ↗
Body morphology, breast tissue, skin characteristics, adhesive contact, motion and clinical access can affect signal quality and usability. Proposed testing should document placement landmarks, adaptations across body types, movement artifacts and any systematic differences in signal quality or adhesion. This is an EasyG design recommendation informed by the statement's placement-error discussion (pp. 1313, 1318), not a claim that the 2007 paper validated EasyG's anatomy-specific approach.
Five foundational engineering pillars governing hardware, firmware, and hospital telemetry integration.
Preserve interpretable ECG morphology across bipolar, augmented, and precordial views.
Minimize weight, bulk, and snag hazards with short coiled leads and a compact sternum pod.
Continuously indicate signal quality, electrode lead-off, battery level, and wireless link loss.
Enable rapid anatomical placement, quick patient pairing, and streamlined hospital disinfection.
Ensure patient electrical isolation/protection and encrypted hospital wireless data transfer.
Structured benchtop verification protocols mapped to each clinical and engineering requirement.
Compare simulated/reference waveforms, amplitude, and morphology against calibrated ECG simulator outputs.
Measure sample rate consistency (≥ 500 SPS), packet timestamps, and dropped samples under sustained streaming.
Assess signal quality index and baseline stability during simulated patient movement and coiled-lead handling.
Confirm immediate detection and central-display indication of disconnected or high-impedance electrodes.
Measure link loss recovery, access-point roaming reconnection, and fail-safe patient ID association.
Evaluate continuous battery runtime, skin adhesive retention, and ergonomic comfort across body types.
A visually plausible simulated waveform is not proof that the electronics, lead calculations or clinical decisions are accurate. These three verification tracks require different evidence.
Inject calibrated ECG simulator signals into the AFE. Record measured bandwidth, input noise, gain/amplitude error, sample rate, channel alignment and filter-induced ST deviation. Log instrument calibration and test conditions.
Confirm identities such as II = I + III under a shared electrode-potential model, then quantify latency, numerical error and waveform changes under noise, lead-off, lead swapping and motion artifacts. Compare filtered versus unfiltered features.
For future appropriately approved studies, compare EasyG with a reference ECG under matched posture, timing and placement documentation. Report waveform amplitudes, intervals, error distributions and pathology-relevant failure modes separately.
If representative/median beats are constructed, document the algorithm and resulting noise; the AHA statement notes approximately 5 µV template noise for resolving 20 µV deflections to 10% error (pp. 1310–1311). Use time-coherent multi-lead measurements for global intervals when available; do not infer global durations from one display view. Label simulated, derived, modified-placement and recorded data distinctly. Physician interpretation and independent validation would be required before diagnostic claims.
Design rationale: Kligfield et al., AHA/ACCF/HRS scientific statement (2007) ↗. Test plans are proposed EasyG work, not completed results.
EasyG began with exploratory research into distributed, wire-free ECG acquisition and mathematical lead reconstruction. Concept selection narrowed the scope to a five-electrode, single-transmitter hospital telemetry system. The current concept retains short electrode wires while replacing the long connection to bedside monitoring equipment with a proposed Wi-Fi link.

Investigated UWB timing, local differential acquisition, electrode geometry and methods that derive conventional leads from a reduced measurement set.
Compared distributed nodes, flexible strips, transmitter clips and a centralized wearable using clinical usability, complexity, signal quality, cost and engineering risk.
Focused on dependable ECG acquisition and wireless transport through a patient-worn ESP32 system, with measured performance and clinical workflow integration still pending.
Lv and colleagues (2026) reported a distinct three-node UWB-synchronized lead-derivation system. Their reported synchronization and waveform agreement outcomes describe their experimental system, not this capstone prototype. EasyG's present five-electrode Wi-Fi design uses a different architecture and must be characterized independently.
The interactive ECG explorer above uses a single set of synthetic electrode potentials to calculate its displayed lead views. This is an educational mathematical model—not a test of an actual ECG analog front end, medical alarm, or Wi-Fi telemetry link.
Select a poster to open the original PNG at native resolution. The diagrams are educational reference illustrations; some depict general ECG systems or earlier multi-node design concepts rather than the selected EasyG prototype.
These references inform the project design and future validation. None should be interpreted as experimental verification of EasyG's current architecture.
FOUNDATIONAL SCIENTIFIC STATEMENT · 2007
Kligfield P, et al. Recommendations for the Standardization and Interpretation of the Electrocardiogram: Part I: The Electrocardiogram and Its Technology. J Am Coll Cardiol. 2007;49(10):1109–1127. A scientific statement, not an evidence-graded guideline. Supports acquisition, filtering, lead-placement and validation rationale; it does not establish EasyG compliance.
Read the statement ↗B. Lv et al., Measurement 282, 122044. Separate three-node architecture using synchronized acquisition and session-specific lead transformation.
Read published article ↗Large public clinical ECG dataset that may support offline algorithm exploration; it is not a substitute for bench characterization of the EasyG acquisition hardware.
Explore PTB-XL ↗Particular requirements concerning the basic safety and essential performance of electrocardiographic monitoring equipment. Compliance has not been assessed.
The site will distinguish numerical design targets from independently measured results, test conditions, instrument calibration and documented pass/fail decisions.
Lock component selection for low-noise AFE, ≥16-bit ADC, ESP32 wireless module, and miniaturized enclosure footprint.
Quantify input-referred noise floor, continuous shift runtime, network drop resilience, and lead-off alarm latency.
Refine skin-contact adhesive ergonomics and validate seamless compatibility with managed hospital Wi-Fi gateways.
Execute comprehensive electrical safety, cybersecurity, and diagnostic waveform verification against clinical benchmarks.